Prior to retrieving data from Forney specimens must be sent to Forney via a manual call. Updates to samples which have been created in Forney will be automatically observed by the integration and updated in ForneyVault. The automated tasks checks for samples that have been updated every 30 minutes.

Specimen post calls can be initiated through the "Export to Forney" button. When initiated, all checked records will be sent to Forney through an API call, once successfully transmitted the Exported flag will be changed to "Yes". Once exported a unique key, created by MetaField, will have been sent to Forney and stored in the MetaField database. This field will not be exposed but will serve as the link for record in ForneyVault and MetaField.

If a specimen which has been exported is re-exported MetaField will update the existing record in ForneyVault.

The following fields are sent to Forney from MetaField

  • Project Number
  • Lab ID
  • Sample Number + Specimen Test Age
  • Height (If specimen size defined in admin)
  • Diameter 1 & 2 (If specimen size defined in admin)
  • Set Number
  • Specimen Number
  • Specified Strength
  • Specified Strength Test Age

...

FieldDescription
Width 1
  • By default, this field will not be populated by the integration. The existing value will be retained or one will have to be entered manually.
  • Field can be configured to pull value from Forney Database. Recommended, if measurements are taken at the time of break and entered in Forney interface.
  • Default value can be defined to apply to all test results that do not have a Width 1 (Length 1 in Forney) measurement.


    Tip

    This field can be configured in Administration>General>Integrations. Documentation can be found at General - Integrations or in the specimen size definition in Admin.


Width 2
  • By default this field will not be populated by the integration. The existing value will be retained or one will have to be entered manually.
  • Field can be configured to pull value from Forney Database. Recommended, if measurements are taken at the time of break and entered in Forney interface.
  • Default value can be defined to apply to all test results that do not have a Width 2 (Width 1 in Forney) measurement. 

    Tip

    This field can be configured in Administration>General>Integrations. Documentation can be found at General - Integrations or in the specimen size definition in Admin.


Load Strength (lbs)
  • MaxLoad value recorded by Forney break machine will be used to populate Load Strength.
Capping Method
  • By default this field will not be populated by the integration.
  • A default capping method, selected from a list of customer defined capping methods, can be applied to all test results that do have a value recorded. 

    Tip

    This field can be configured in Administration>General>Integrations. Documentation can be found at General - Integrations.


Is Cut
  • By default this field will not be populated by the integration.
  • A default Is Cut can be applied to all test results when retrieving data from Forney. Otherwise, the value will default to "No".

    Tip

    This field can be configured in Administration>General>Integrations. Documentation can be found at General - Integrations.


Tested Date
  • Test date from Forney will be retrieved and converted to the project's time zone. If a project timezone is not defined the default customer timezone will be used.
Tested By
  • Forney is unable to return enough data to indicate the MetaField user that conducted the test.
  • The integration will retrieve the initials of the lab technician that is logged into the Forney break machine at the time of the test.

Image Added


Prisms

The "Retrieve From Forney" action is available on grout and mortar prism sample types within the To Test queue in LIMS. When the call to Forney is initiated, MetaField will attempt to retrieve break data from Forney for each specimen that appears on the screen. MetaField will populate all data that is present either in the ForneyVault database or through predefined default values in the Administration area.

FieldDescription
Width 1,2,3,4
  • By default, this field will not be populated by the integration. The existing value will be retained or one will have to be entered manually.
  • Field can be configured to pull value from Forney Database. Recommended, if measurements are taken at the time of break and entered in Forney interface.





Height 1,2,3,4
  • By default this field will not be populated by the integration. The existing value will be retained or one will have to be entered manually.
  • Field can be configured to pull value from Forney Database. Recommended, if measurements are taken at the time of break and entered in Forney interface.

Load Strength (lbs)
  • MaxLoad value recorded by Forney break machine will be used to populate Load Strength.

Capping Method
  • By default this field will not be populated by the integration.
  • A default capping method, selected from a list of customer defined capping methods, can be applied to all test results that do have a value recorded. 

    Tip

    This field can be configured in Administration>General>Integrations. Documentation can be found at General - Integrations.



Is Cut
  • By default this field will not be populated by the integration.
  • A default Is Cut can be applied to all test results when retrieving data from Forney. Otherwise, the value will default to "No".


Tested Date
  • Test date from Forney will be retrieved and converted to the project's time zone. If a project timezone is not defined the default customer timezone will be used.

Tested By
  • Forney is unable to return enough data to indicate the MetaField user that conducted the test.
  • The integration will retrieve the initials of the lab technician that is logged into the Forney break machine at the time of the test.

Image Modified